Access 2 Immunoassay System
A COMPACT DESIGN TO FIT YOUR SPACE AND TIME
Beckman Coulter
Analytical system | Fully automated, random-access immunoassay bench top analyzer for low-volume laboratories |
Measurement principles | Dioxetane-based chemiluminescent detection with magnetic particle separation |
Analytical method | Chemiluminescent detector: Luminometer |
Throughput | 100 tests/hr maximum |
Compartment Temperatures | Incubator and wash/read carousel: 37°C Sample carousel: 4 – 10°C
Reagent compartment: 4 – 10°C |
Sampler capacity |
Holds up to 24 self-sealing reagent packs |
Chemistry menu capacity | > 50 preprogrammed, barcoded immunoassay methods currently available** |
Barcoded Reagents | Automatic tracking of:
Number of tests Available tests Expiration date Lot number Calibration expiration |
Calibration | Calibration curve stability up to 56 days (assay dependent) Calibration curves and parameters displayed on screen and printouts |
Sample barcode formats | Sample barcode formats Code 39 (Code 3 of 9 or SD-3) Code 128 (USD-6) Interleaved 2 of 5 (USD-1) Codabar (USD-4) Sample volume 5 to 200 μL (assay dependent) < 50 μL is typical |
Sample container sizes | Primary tubes: 12, 13 x 75 mm glass or plastic 16 x 75 mm 13, 16 x 100 mm Sample cups: 2.0 mL, 3.0 mL 1.0 mL, 2.0 mL insert cups Pediatric insert Auto aliquot tube |